ISO 10993Biological evaluation of medical devices
The international standard for biological evaluation of medical devices — the risk-based programme that decides which biological endpoints (cytotoxicity, sensitisation, irritation, systemic toxicity, genotoxicity, implantation, chemical characterisation) must be tested for a given patient-contact category.
01What ISO 10993 actually is
ISO 10993 is a 23-part international standard family (10993-1 through 10993-23) covering the biological evaluation of medical devices. Part 1 — ISO 10993-1:2018 — is the framework standard: it sets out the risk-management approach to biocompatibility, classifies devices by nature and duration of patient contact, and drives the decision about which biological endpoints (Parts 3 through 23) must be evaluated through test, literature, justification or chemical characterisation.
FDA recognises ISO 10993-1:2018 as a consensus standard and issued a 2023 guidance giving the agency's specific expectations. EU MDR Annex I §10 makes biocompatibility a General Safety and Performance Requirement, and the harmonised version of ISO 10993-1 is the route to GSPR conformity. Health Canada, PMDA, TGA, ANVISA and essentially every other regulator follow the same framework.
02Device categorisation — the matrix that drives the programme
10993-1 categorises devices on two axes: nature of body contact and duration of contact.
| Nature of contact | Examples |
|---|---|
| Surface — intact skin | Stethoscopes, electrodes. |
| Surface — mucosal membrane | Contact lenses, urinary catheters. |
| Surface — breached / compromised surface | Wound dressings, ulcer dressings. |
| External communicating — blood path indirect | Solution-administration sets, blood-collection sets. |
| External communicating — tissue / bone / dentin | Laparoscopes, dental cements. |
| External communicating — blood path direct | Catheters intravascular, dialysers. |
| Implant — tissue / bone | Orthopaedic implants, sutures, intraocular lenses. |
| Implant — blood | Stents, prosthetic heart valves, vascular grafts. |
| Duration of contact | Definition |
|---|---|
| Limited (A) | ≤24 hours, single or multiple use cumulative. |
| Prolonged (B) | Greater than 24 hours, ≤30 days. |
| Long-term (C) | Greater than 30 days. |
The intersection of these two axes produces a cell in Annex A Table A.1 of 10993-1, which lists the biological-effect endpoints to be evaluated — typically eight to twelve endpoints for a long-term implant and one or two for a transient intact-skin contact.
03The biological endpoints
| Endpoint | ISO 10993 part | When required |
|---|---|---|
| Cytotoxicity | Part 5 | Almost every device. |
| Sensitisation | Part 10 | Almost every device — Magnusson-Kligman, LLNA or human repeat-insult patch test. |
| Irritation / Intracutaneous reactivity | Part 23 (replaces Part 10 sections) | Almost every device. |
| Acute / sub-acute / sub-chronic systemic toxicity | Part 11 | Implants and external-communicating > 24 h. |
| Material-mediated pyrogenicity | Part 11 | Blood-contact and implants. |
| Genotoxicity | Part 3 | Permanent implants, long-term external-communicating. |
| Carcinogenicity | Part 3 | Permanent implants where genotoxicity is positive or material is novel. |
| Reproductive / developmental toxicity | Part 3 | Long-term implants intended for reproductive contact. |
| Implantation | Part 6 | All implants. |
| Haemocompatibility | Part 4 | Blood-contact devices. |
| Chemical characterisation | Part 18 | Foundation for the toxicological risk assessment in modern submissions. |
| Toxicological risk assessment | Part 17 | Synthesises chemical characterisation into a risk decision. |
04Chemical characterisation first — the modern programme
ISO 10993-18:2020 reset the discipline. Rather than reflexively running every animal test on every endpoint, modern programmes start with chemical characterisation: identify the materials of construction, the manufacturing process, the cleaning and sterilisation regime, then perform extractables and leachables (E&L) studies under exaggerated and simulated-use conditions to identify and quantify every compound that could leach into the patient.
The extractables list goes to a toxicological risk assessment (ISO 10993-17:2023) where each compound is evaluated against published toxicity data and a calculated Acceptable Exposure (AE) is compared to estimated patient exposure. Endpoints where the toxicological risk assessment shows margin can be addressed without biological testing — directly satisfying the 3Rs (replacement, reduction, refinement of animal testing) and the EU MDR Annex I §10.4 hierarchy of evidence.
05The Biological Evaluation Plan and Report
10993-1 §5 requires a Biological Evaluation Plan (BEP) and Biological Evaluation Report (BER). The BEP is written before testing starts: it captures the device categorisation, the endpoints identified, how each endpoint will be addressed (test, literature, characterisation + risk assessment, justified absence), and the test methodologies and acceptance criteria. The BER is the closeout: it presents the evidence against each endpoint and concludes on biological safety.
The BER is the document the Notified Body or FDA reviewer reads. It must be self-contained — every endpoint addressed with traceable evidence, every justification scientifically defensible, and the conclusion explicitly linking the biological evaluation to the device's intended use and the ISO 14971 risk file.
06Common Notified Body / FDA findings
- Device categorisation wrong — sites classify a blood-path device as external-communicating-indirect when it is direct, missing the haemocompatibility programme.
- Biological evaluation copied from a predecessor device without justification that manufacturing process, materials, sterilisation and shelf life are equivalent.
- Chemical characterisation (Part 18) extractables study run under conditions less aggressive than worst-case clinical use.
- Toxicological risk assessment (Part 17) AE values derived from non-pharmacopoeial sources without traceability.
- Sensitisation addressed with LLNA only — FDA increasingly expects guinea pig maximisation for skin-contact polymeric materials.
- Sterilisation residuals (EtO, ECH) not addressed under ISO 10993-7 when EtO is the chosen sterilisation modality.
- Particulate matter from polymer-on-polymer wear addressed by literature only, no Part 6 implantation study with histopathology.
- BER concludes biological safety but the underlying chemical-characterisation report has gaps the BER does not acknowledge.
07How V5 Ultimate is built around ISO 10993 design history
- BEP template captures categorisation, endpoint matrix and test plan, with cross-links to the ISO 14971 risk file.
- Material declarations are first-class objects — change to a polymer compound triggers a biocompatibility-re-evaluation ticket.
- Sterilisation modality is tracked alongside the device record so EtO devices automatically flag the -7 residuals programme.
- Test reports from contract labs (Toxikon, NAMSA, WuXi AppTec, Eurofins) are filed against their endpoint, with expiry-based re-test reminders.
- BER drafts pull every endpoint conclusion from the same dossier so the closeout document cannot lose synch with the underlying reports.
- Chemical characterisation (Part 18) and toxicological risk assessment (Part 17) outputs are versioned and linked to the analytical-method qualifications.
Frequently asked questions
Q.Is ISO 10993-1:2018 mandatory?+
FDA recognises ISO 10993-1:2018 as a consensus standard and the 2023 guidance is the agency's stated expectation. EU MDR Annex I §10 makes biocompatibility a GSPR. Health Canada, PMDA, TGA, ANVISA, MFDS and other regulators follow the same framework. Departures are permitted with justification but in practice every patient-contacting device follows 10993-1.
Q.Can I avoid animal testing entirely?+
Increasingly yes, for many endpoints. ISO 10993-18 chemical characterisation plus ISO 10993-17 toxicological risk assessment can replace acute/sub-acute systemic toxicity, irritation and even sensitisation in some submissions. Implantation, haemocompatibility and certain genotoxicity endpoints still typically require biological testing. The EU MDR and FDA both encourage the 3Rs hierarchy.
Q.Does an equivalent predecessor device avoid biological testing?+
Sometimes. ISO 10993-1 §5.2 allows leveraging prior data if equivalence of materials, manufacturing process, sterilisation, shelf life and intended use is rigorously demonstrated. FDA Special 510(k) and EU MDR equivalence claims accept this but require a documented gap analysis — not assertion.
Q.How long are biocompatibility test reports valid?+
The reports do not expire on a calendar. They remain valid as long as the device, materials, manufacturing process, sterilisation regime, shelf life, packaging and intended use are unchanged. Any change triggers an ISO 14971 / change-control evaluation and may trigger re-test.
Q.What is the difference between ISO 10993-7 and ISO 10993-10?+
10993-7 covers ethylene oxide (EtO) sterilisation residuals — the EtO and ECH limits in the finished device. 10993-10 historically covered sensitisation and irritation, but the irritation content has been moved to the new ISO 10993-23:2021. Sensitisation methods (Magnusson-Kligman, LLNA, HRIPT) remain in 10993-10.
Primary sources
Further reading
- ISO 14971 — Risk management10993-1 §4 makes 14971 the framework for the biocompatibility risk decision.
- IEC 60601 — Medical electrical equipment60601-1 §11.7 references 10993 for applied-part biocompatibility.
- ISO 13485 — QMSHolds the design-control evidence the biocompatibility file lives inside.
- Design controls (21 CFR 820.30)US framework that requires biocompatibility evidence in the DHF.
Explore this topic
ISO 10993 sits inside this topic cluster in our glossary. Every neighbour is one click away.
Device-specific rules, submissions and the standards that bind them.
V5 Ultimate ships with the ISO 10993 controls already wired in — audit trail, e-signatures, validation evidence. Free trial, no credit card, onboard in days, not months.
